Applicable Syringes | 10ml, 20ml, 30ml, 50ml, 5ml, 60ml of any standard |
---|
Accuracy | +/-2% with correct calibration |
---|
Rate Mode | 0.1ml/h-2000.0ml/h (0.1ml/h or 1ml/h step) |
---|
Time Mode | SOL. VOL: 0.1-9999.9ml (0.1ml step) |
---|
Dose Mode | Volume: 0.1-999.9ml (0.1ml step) |
---|
Total Volume Range | 0.1-9999.9ml (0.1ml step) |
---|
Drug Library Mode | Available (only for SP3) |
---|
Sequential Mode | Available (only for SP3) |
---|
Purge Rate | Programmable and adjustable |
---|
KVO Rate | 0.0~5.0 ml/h, Programmable and adjustable |
---|
DPS Occlusion Pressure | LOW: 60(+15)kPa |
---|
Alarms | Empty, Infusion Complete, Loose, Low Battery, Near Empty, No Action, No Battery, Occlusion, Rate Abnormal, Syringe Plunger Disengagement |
---|
Alarm Volume | MEDIUM or HIGH, Three levels: LOW |
---|
Bolus Rate | 0.1-2000ml/h, Automatic/Manual bolus available, Programmable and adjustable |
---|
Bolus Volume | Range: 1.0~50.0ml (0.1ml step) |
---|
Interface | RS-232 |
---|
Brightness | 1~10 levels can be selected |
---|